Fluke 8842a Portable Multimedia Player User Manual


 
Maintenance
TROUBLESHOOTING
6
6-43
Configures the 8842A in the mA DC function and the slow reading rate, and measures
the reading across the 0.1 current shunt. This test should be fairly immune to outside
noise because the total driving impedance is typically less than 1 k. The reading is not a
perfect zero because of the offsets generated by charge injection of U302 and the T/H
Amplifier (X10 configuration).
TEST 5: 200 VDC, Zero
Configures the 8842A in the 200V dc range and slow reading rate. Input noise is divided
by 100. Assuming no input of any kind, the T/H Amplifier is essentially shorted to
ground by 100 k and filtered by the 3-pole analog filter. Any non-zero reading under
quiet input conditions is due to the offset of the T/H Amplifier (X1 configuration).
TEST 6: 1000 VDC, Zero
Configures the 8842A exactly as in the 1000V dc range and slow reading rate, with input
noise being divided by 1000. The reading is very close to zero because of the inherent 2
k driving impedance to the T/H Amplifier (X1 configuration).
TEST 7: 1000 VDC + 20 M
Couples the 1000V dc range and 20 M current source together. The result is nominally
500 nA through the 10 M input divider. Since the 1000V dc range senses this voltage at
the divide-by-1000 point of the scaling circuit, the reading should be 5 mV, or 500 counts
at the A/D Converter. This test could indicate an error if input capacitance is greater than
1000 pF.
TEST 8: 20 VDC + 20 M
Puts the DC Scaling circuit into the 20V dc range and the Ohms Current Source into the
20 M range. The infinite input impedance of the 20V dc range causes the 20 M
current source to be clamped at its maximum open circuit voltage, typically 12V. The
20V dc range scales this voltage and presents the A/D Converter with 1.15V, or 115,000
counts.This is a good test to ensure that the maximum open-circuit voltage of the Ohms
Current Source is less than 13V. This test is susceptible to capacitance greater than 0.01
µF at the input terminals.
TEST 9: 20 VDC + 2000 k
Puts the DC Scaling circuit in the 20V dc range and the Ohms Current Source in the 2000
k range. The infinite input impedance of the 20V dc range causes the 2000 k current
source to be clamped at typically 11.5V. The reading at the A/D Converter should be
1.15V. Again tests that the maximum open-circuit voltage of the Ohms Current Source is
less than 13V. Capacitances greater than 0.1 µF at the input terminals can cause an error.
TEST 10: 2 VDC + 2000 k
Puts the DC Scaling circuit in the 2V dc range and the Ohms Current Source in the 2000
k range, except that its maximum open-circuit voltage is limited to less than 6.5V in
this configuration. This test, as well as tests 11-13, checks clamps Q312 and Q313 and
the analog filter. The reading at the A/D Converter should be an overload. Capacitances
greater than 0.5 µF at the input terminals can cause an error.
TEST 11: 200, Overrange
TEST 12: 2 k, Overrange
TEST 13: 20 k, Overrange
TEST 14: 200 k, Overrange