Description of Control Keys:
■ The CH1 VOL and CH1 ZERO keys for CH1 are used to set the horizontal
scale and position.
■ The CH2 VOL and CH2 ZERO keys for CH2 are used to set the vertical scale
and position continuously.
The following functions are not available in the XY display mode:
■ Reference or digital value waveform
■ Cursor
■ Time base control
■ Trigger control
STORINGWAVEFORMS
The oscilloscope can save four (4) waveforms that can be displayed on the screen along
with the present waveform. The recalled waveforms cannot be adjusted.
The waveform Store / Recall menu is described in the following Table.
Function
menu
Setups Description
Source CH1
CH2
MATH
Select the signal source of the waveform to save
(the waveform to save must be on the display)
WAVE A, B, C and
D
Select the memory address for saving or recalling
data
Save Store the waveform of the selected signal source
into the selected address
Show ON
OFF
Set ON to display the waveforms stored in address
A, B, C or D. Choose OFF to exit this feature.
To save a CH 1 waveform in address A, perform the following steps:
1. Press the MENU key; the function menu will appear at the right of the screen.