45
V5.1‐March2010
ACQUISITIONMODE
The Acquiring Mode menu is described in the Table shown below.
Function menu Settings
Description
Sample The waveform data sampling time interval. The
sample mode accurately reconstructs the waveform,
but cannot respond to rapid changes and sudden
peaks.
Peak Detect 50ns
max.
The peak detect mode captures rapid changes and
sudden peaks.
Average Multiple samples are averaged. The average mode
reduces the noise level (for best results, the waveform
must be repetitive)
Averages 4, 16, 64
or 128
Select the number of samples to average
DISPLAYSETTING
The Display Setting menu is described in the following table.
Function menu
Settings
Description
Type Vectors
Dots
Shows the waveform as a smooth line, connecting each
data point
Shows the waveform as a collection of independent
data points
Persist OFF
1s
2s
5s
Infinite
Sets the duration of time that previous waveforms
remain on the display (useful for observing waveform
variations)
Format YT
Displays the relative relationship between vertical
voltage and horizontal time.
Displays CH1 on the horizontal axis and CH2 on the