A SERVICE OF

logo

DCP42enGB_V1.512/14
14
6.5.3RemoteControlOutputON/OFF
RemoteoutputON/OFFcontrolcanbeactivatedinanyofthemodesNormal,Preset,Set,and
Remote.Reviewtheconditionsbelow:
Bydefault,Pin5isopenandtheoutputisON.
ShortingPin5toPin4(ground)switchestheoutputOFF.
WhentheoutputisOFF,theC.V.&C.C.LEDswillflash. Thepresentoutputvoltageand
currentsettingwillbeshownonthepanelmeter.
Theusercanadjusttheoutputusingthevoltage&currentcontrolknobtothedesired
valuewhentheoutputisOFF.
Remark:Useonlythe8pinremoteplugprovidedandconnectwith22AWGwires.See
diagrambelow.
Note: Pin numbers are marked on the black area of the socket
7. PCSoftwareProgram
ThismeterhasthecapabilitytoconnecttoandcommunicatewithaPC.Toinstallandusethe
software,pleaserefertotheinstructionsprovidedonthesuppliedCDROMand/ortheinstructions
providedintheHELPUtilitywithinthesoftwareprogram.
Checkthesoftwaredownloadpageofthewebsitewww.extech.comforthelatestversionofthePC
softwareanditsoperatingsystemcompatibility.
8. AutomaticOutputStepPrograms
TousetheAutomaticOutputSteputilitytheusermustfirstbecomefamiliarwiththePC
interfaceandthesuppliedsoftware.Toinstallandusethesoftware,pleaserefertothe
instructionsprovidedonthesuppliedCDROMand/ortheinstructionsprovidedintheHELP
Utilitywithinthesoftwareprogram.Oncefamiliaritywiththesoftwareisachievedrefertothe
instructionsbelow:
TheunitcanbeexternallyprogrammedviaaPCtorun20steps,eachstepwithapreset
voltage/currentlimitingvalueandapresettimeperiodof1secondto99minutes.Thetimed
programcanbesetto
runonecycleoruptoaninfinitenumberofcycles.