6.5RemoteControlOperation
Therearetwomethodsforremotelycontrollingcurrentandvoltageadjustments.Currentmust
becontrolledforbothmethodsotherwisetheunitwilldefaulttoCCmode.
6.5.1REMOTECONTROL–METHOD1:UsingtwoexternalDCvoltagesources
RemoteSocketPinAssignmentforexternalvariablevoltagesource
PIN FUNCTIONS REMARKS
1 InternalDC+5V Lowerthan50mA
2 Voltageadjust 0to5V
3 Currentadjust 0to5V
4 Ground
5 OutputOFF Shorttoground
6 N.A.
7 N.A.
8 N.A.
Checktheoutputvoltagerangeofthepowersupplybyvaryingtheexternalvoltagesource.
Shortcircuitthemainoutputwith14AWGwiretochecktheCCsettingonthe displaywhile
varyingtheexternalvoltagesource.
6.5.2REMOTECONTROL–METHOD2:Usingtwo0to5KOhm
variableresistors
RemoteSocketPinAssignmentforexternalvariableresistor
PIN FUNCTIONS REMARKS
1 InternalDC+5V Oneendofresistor
2 Voltageadjust Variablelegofresistor
3 Currentadjust Variablelegofresistor
4 Ground Otherendofresistor
5 OutputOFF Shorttoground
6 N.A.
7 N.A.
8 N.A.
Checktheoutputvoltagerangeofthepowersupplybyadjustingthe5Kohmvariableresistor.
Shortcircuitthemainoutputwith14AWGwiretochecktheCCsettingonthe displaywhile
adjustingthevariableresistor.