6.3UpperVoltageandCurrentLimits
Whenthevoltageorcurrentattheoutputterminalexceedsthemaximumallowablelimits,the
outputwillautomaticallyswitchOFF.Thisadditionalprotectionfeatureisnecessaryforapower
supplywithsuchawiderangeofvoltageandcurrentcapabilities.
TosettheUVL:
1. TosettheUVL:PresstheSHIFTbuttonandthenpresstheVoltagecontrolknob.The
voltmeterwillshowthecurrentlimitsettingandtheAmmeterwillshowSuul.
2. TurntheVoltagecontrolknobtothedesiredlimitvalue.PressSHIFTtoconfirmandto
returntonormaloperation.
3. TosettheUCL,pressSHIFTandthenpresstheCurrentcontrolknob.TheAmmeterwill
showthecurrentlimitsettingandtheVoltmeterwillshowSUCL.
4. TurntheCurrentcontrolknobtothedesiredlimitvalue.PressSHIFTtoconfirmandto
returntonormaloperation.
Note:
TheUVLandUCLsettingsdefaulttothemaximumoutputvalue.
6.4RemoteSensing
Note:ObservewarningsdisplayedbythePowerSupply,anincorrectdisconnectionsequence
willdamagethePowerSupply
Warnings:NevershorttheRemoteSensingTerminaland alwaysdisconnecttheRemoteSensing
Terminalfirst.
6.4.1Connection
1.Firstcompletethepowerconnectionsbetweenpowersupplyandequipment.
2.Checkandensurethatthe
powerconnectionsaresecure.
3.MakeconnectionsbetweenRemoteSensingandequipment.
Warning!:NevershorttheRemoteSensingTerminaland neverconnecttheRemoteSensing
Terminalinreversepolarity
ThediagrambelowshowstheconnectionsforRemoteSensing,Poweroutput,andEquipment.
TheremotesensingwireshouldbeATLEAST
22AWGwiresize.
6.4.2Disconnection
Warning!Incorrectdisconnectsequenceswilldamagethepowersupply
1.Firstdisconnecttheremotesensingconnections.
2.Thendisconnectthepowerconnectionsbetweenthepowersupplyandequipment.