© National Instruments Corporation 5 NI PXI-5600 Calibration Procedure
Test Conditions
Follow these guidelines to optimize the connections and the environment
during calibration:
• Keep connections to the NI PXI-5600 short. Long cables and wires act
as antennae, picking up noise that can affect measurements.
•Use 50 Ω coaxial cable for all connections to the NI PXI-5600.
• Keep relative humidity between 10 and 90%, noncondensing,
or consult the hardware documentation for the optimum relative
humidity.
Documentation
The following documents contain additional information you might find
helpful when calibrating the NI PXI-5600:
• NI-TUNER Reference Help at Start»Programs»National
Instruments»RF Signal Analyzer»Documentation»Low Level
• PXI-5660 RF Signal Analyzer Specifications at Start»Programs»
National Instruments»RF Signal Analyzer»Documentation
• PXI-5660 RF Signal Analyzer User Manual at Start»Programs»
National Instruments»RF Signal Analyzer»Documentation
• Spectral Measurements Toolset User Guide at Start»Programs»
National Instruments»RF Signal Analyzer»Documentation
Broadband
Precision Power
Splitter
Weinschel 1507R Frequency Range: DC to 3.0 GHz
Output Maximum
Voltage Standing-Wave
Ratio (VSWR):
1.15
Input Maximum VSWR: 1.25
Amplitude Tracking: <0.1 dB
2
Phase Tracking: <4 Deg
Insertion Loss: Nominal 6 dB
1 Including splitter and adapter uncertainty: 0.11 dB standard uncertainty when unit under test (UUT) on other side of the
splitter has SWR of up to 2.5.
2 Screened for
<0.1 dB tracking.
Table 1. Required Test Instruments (Continued)
Required
Equipment
Recommended
Make and Model
Necessary Specifications