NI PXI-5600 Calibration Procedure 4 ni.com
Table 1. Required Test Instruments
Required
Equipment
Recommended
Make and Model
Necessary Specifications
Spectrum Analyzer Agilent 8563E Frequency Range: 109 MHz to 111 MHz
Noise Floor: <130 dBm/Hz
Frequency Resolution: >0.5 Hz at 110 MHz
Lockable to External Frequency Reference
Signal Generator Agilent
ESG-4422B
Frequency Range: 20 MHz to 2.8 GHz
Amplitude Range: –30 dBm to +16 dBm
Phase Noise/
Spectral Purity:
<95 dBc/Hz at 1 KHz
Power Meter with
2 Power Sensors
Anritsu
ML2438A
MA2421A
MA2473A
Range of IF Output
Power Sensor on CH-A:
–30 dBm to +20 dBm
Range of Frequency of
Sensor on CH-A:
3MHz to 27MHz
Absolute Accuracy
(Standard Uncertainty):
0.07 dB
Range of RF Input Power
Sensor on CH-B:
–40 dBm to +20 dBm
Range of Frequency of
Sensor on CH-B:
20 MHz to 2.8 GHz
Absolute Accuracy
1
(Standard Uncertainty):
0.11 dB
Frequency
Reference
Datum 8040
Rubidium
Frequency
Standard
Frequency: 10 MHz
Frequency Accuracy: 1 ppb (typically ±6E-10
within 1 year)
Amplitude Range: >5 dBm (typ: 7 dBm)