24
DL-00096 Rev. 7
Immunity test
Voltagedips,short
interruptions, and
voltagevariations
on power supply
input lines
IEC61000-4-11
Powerfrequency
(50/60Hz)magnetic
field
IEC61000-4-8
ConductedRF
IEC61000-4-6
RadiatedRF
IEC61000-4-3
IEC 60601 test
level
<5%U
T
(>95%dipinU
T
)
for0.5cycle
40%U
T
(60%dipinU
T
)
for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dipinU
T
)
for5sec
3A/m
3Vrms150kHz
to80Mhz
20V/m80Mhz
to2.5Ghz
Electromagnetic
environment/guidance
Mainspowerqualityshould
be that of a typical
commercial or hospital
environment.Iftheuser
ofthePortableOxygen
Concentratorrequired
continued operation during
power main interruptions,
it is recommended that the
concentrator be powered
from an uninterruptible
powersupplyorbattery.
Powerfrequencymagnetic
fieldsshouldbeatlevels
characteristic of a typical
commercial or hospital
environment.
PortableandmobileRF
communications equipment
should be used no closer
toanypartofthePortable
OxygenConcentrator,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the
transmitter.
Recommended separation
distance:
d=1.2√P
d=1.2√P80MHzto800
MHz
d=2.3√P800MHzto2.5
GHz
wherePisthemaximum
output power rating of the
transmitterinwatts(W)
according to the transmitter
manufacturer and d is the
recommended separation
distanceinmeters(m).
Fieldstrengthsfromfixed
RFtransmitters,asdeter-
mined by an electromagnetic
sitesurvey
a
should be less
thanthecompliancelevel
in each frequency range
b
.
Interferencemayoccurinthe
vicinityofequipmentmarked
with the following symbol:
Compliance
level
<5%U
T
(>95%dipinU
T
)
for0.5cycle
40%U
T
(60%dipinU
T
)
for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dipinU
T
)
for5sec
3A/m
3Vrms
20V/m
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
a
Fieldstrengthsfromfixedtransmitters,suchasbasestationsforradio
(cellular/cordless)telephonesandlandmobileradios,amateurradio,AM
andFMradiobroadcast,andTVbroadcastcannotbepredictedtheoretical-
lywithaccuracy.Toassesstheelectromagneticenvironmentduetofixed
RFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.If
themeasuredfieldstrengthinthelocationinwhichthePortableOxygen
ConcentratorisusedexceedstheapplicableRFcompliancelevelabove,
theconcentratorshouldbeobservedtoverifynormaloperation.Ifabnormal
performanceisobserved,additionalmeasuresmaybenecessary,suchas
reorientingorrelocatingtheconcentrator.
b
Overthefrequencyrange150kHzto80MHz,fieldstrengthsshouldbe
lessthan3V/m.
DL-00096 rev 7.indd 24 6/1/2009 4:48:35 PM