Serial Communications 460L Instruction Manual
36 05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED DCN 5164
7.3.8. TLIST
SYNTAX
<address>TLIST<CR>
DESCRIPTION
Returns a formatted, easy to read list of the parameters by name with the current values for
each.
DATA PARAMETERS SENT
None
RESPONSE
Test Parameter List (See Below)
EXAMPLE
Command:
1TLIST<CR>
Response:
O3 = 0.0226168
Press = 14.7753
Cell Temp = 300.7116
Lamp Temp = 324.7965
Ref = 2881.52
Meas = 2941.092
Raw Ref = 4412.646
HI Alarm = ON
HI-HI Alarm = OFF
NOTES
The set of parameters is not the same as those returned by the TDUMP command.
While the concentration value returned shows more digits after the decimal than the front panel
display (in the example above, 0.0226168), it is only valid to 4 significant digits (in the example
above, 0.022).
No checksum is sent in the response to the TLIST command.