The function of each of the default Data Channels is described below:
CONC:
Samples NO
X
, NO and NO
2
(NO
X
, NO and O
2
if the instrument is equipped
with the Oxygen option) concentration data at one minute intervals and stores
an average every hour with a time and date stamp. Readings during calibration
and calibration hold off are not included in the data. The last 800 hourly
averages are stored.
PNUMTC:
Collects sample flow and sample pressure data at five minute intervals and
stores an average once a day with a time and date stamp. This data is useful for
monitoring the condition of the pump and critical flow orifice (sample flow)
and the sample filter (clogging indicated by a drop in sample pressure) over
time to predict when maintenance will be required. The last 360 daily averages
(about
1 year) are stored.
CALDAT:
Logs new slope and offset every time a zero or span calibration is performed,
also records the sample concentration reading just prior to performing a
calibration.
NOTE:
This Data Channel collects data based on an event (a calibration) rather than a timer.
This Data Channel will store data from the last 200 calibrations. This does not
represent any specific length of time since it is dependent on how often calibrations are
performed. As with all Data Channels, a time and date stamp is recorded for every data
point logged.
Table 5-5: DAS Data Channel Editing
Step Action Comment
1.
Press SETUP-DAS-EDIT
Enter DAS menu to edit Data Channels
2.
Press PREV/NEXT
Select Data Channel to edit
3.
Press EDIT
Enter the Edit menu for the selected Data Channel
4.
Press SET> (5 times)
Scroll through Data Channel properties until RS-232
REPORT: OFF is displayed
5.
Press EDIT
Edit selected setup property
6.
Toggle OFF to ON
Change RS-232 REPORT property
7.
Press ENTR
Accepts change
8.
Press EXIT (4 times)
Exits back to sample menu