Fluke 433/434
Getting Started
12
TREND screen.
1
2
The Trend screen shows the changes over time of
measuring values on one row of the table. An example is
the VOLTS/AMPS/HERTZ TREND. Time is displayed
horizontally. The display is built up from the right side of
the screen. To allow continuous recording of data, the
time axis is compressed when necessary.
Screen information:
c
Present values of the trend graphs on
the right side of the screen. If CURSOR
is ON, the trend values at the cursor
are displayed.
d
Trend display area.
Function keys:
F1
Assigns up/down arrow keys to select a
row from the table to be displayed as a
trend. The selected row is indicated in
screen header.
F2
CURSOR ON/OFF.
F3
Assigns the arrow keys to CURSOR or
ZOOM operation. Moving the cursor
across the extreme left or right of the
screen makes the next screen out of a
maximum of 6 visible. ZOOM allows
you to expand or shrink the display to
view details or to see the complete
graph within the screen area.
F4
Return to previous screen.
F5
Switches between HOLD and RUN.