Function Range Resolution Accuracy
Frequency
5.000Hz 0.001Hz
(1.5% reading + 5 digits)
50.00Hz 0.01Hz
500.0Hz 0.1Hz
(1.2% reading + 3 digits)
5.000kHz 0.001kHz
50.00kHz 0.01kHz
500.0kHz 0.1kHz
5.000MHz 0.001MHz
(1.5% reading + 4 digits)
Sensitivity: >0.5V rms min. 1MHz; >3V rms > 1MHz
NOTE:
Accuracy is stated at 18
o
C to 28
o
C (65
o
F to 83
o
F) and less than 75% RH.
Diode Test Test current of 0.3mA typical, open circuit voltage 1.5V DC typical
Continuity Check Audible signal will sound if the resistance is less than 150 (approx.), test
current <0.3mA
Temperature Sensor Requires type K thermocouple
Input Impedance 7.8M
AC Response Average responding
ACV Bandwidth 50Hz to 400Hz
Display 4,000 count liquid crystal
Overrange indicationOL” is displayed
Auto Power Off 30 minutes (approximately)
Polarity Automatic (no indication for positive); Minus (-) sign for negative
Measurement Rate 2 times per second, nominal
Low Battery Indication
is displayed if battery voltage drops below operating voltage
Battery Two AAA batteries
Fuses mA, µA ranges; 0.5A/660V ceramic fast blow
A range; 10A/250V ceramic fast blow
Operating Temperature 5C to 40C (1F to 104F)
Storage Temperature -20C to 60C (-4F to 140F)
Operating Humidity Max 80% up to 87F (31C) decreasing linearly to 50% at 40C (104F)
Storage Humidity <80%
Operating Altitude 2000meters (7000ft.) maximum.
Weight 225g (0.5lb) (includes holster).
Size 135 x 68 x 43mm (5.31” x 2.67” x 1.7”)
Safety This meter is intended for indoor use and protected, against the users, by
double insulation per EN61010-1 and IEC61010-1 2nd Edition (2001) to
CAT II 1000V & CAT III 600V; Pollution Degree 2.
Approvals CE