nition of the buffer press the “ON/OFF” but-
ton to return to normal mode.
For a two point pH calibration, place the
probe in pH 7.01 buffer (or pH 6.86, if the
NIST buffer set was selected).
The meter will recognize the
buffer value and then display
pH 4.01 USE.
Place the probe in the second
buffer (pH 4.01 or 10.01, or, if
using NIST, pH 4.01 or 9.18).
When the second buffer is rec-
ognized, the LCD will display
“OK” for 1 second and the
meter will return to normal mode.
The “CAL” tag is displayed on the LCD to
indicate that the meter is calibrated.
EC/TDS OPERATIONS:
The user can select the EC/TDS conversion
factor (CONV) value, as well as the β coef-
ficient (BETA) for automatic temperature com-
pensation.
Enter the EC or TDS mode with the “RANGE”
key, then press and hold the “ON/OFF” key
until “TEMP” is displayed on the
secondary LCD. Press this key
again to show the current con-
version factor (CONV).
Press the “SET” button to
change the value, then press
again this key to show the cur-
rent temperature coefficient
(BETA).
Press the “SET” button to change the value,
then press the “ON/OFF” key return to normal
operation.
For taking EC or TDS measurements, immerse
the probe in the solution to be tested.
Select either EC or TDS mode with the
“RANGE” key.
Stir gently and wait for the reading to stabi-
lize (the clock symbol stops blinking).
EC/TDS CALIBRATION PROCEDURE:
Clean the probe with M10000 solution.
Enter the EC mode with the “RANGE” key.
Immerse the probe in MA9030 (12.88 mS/cm)
calibration solution.
Press and hold the “ON/OFF”
key until “CAL” is displayed on
the LCD. Release the button
and the message “12.88 mS
USE” will appear.
The meter will perform automatic calibration,
and then display “OK” for 1 second before
returning to normal mode.
The calibration is now complete and the “CAL”
tag is displayed on the LCD to indicate that the
meter is calibrated.
BATTERY REPLACEMENT:
The meter shows the remaining battery percent-
age every time it is switched on. When the
battery level is below 5%, the battery symbol on
the bottom left of the LCD lights up to indicate a
low battery condition.
The meter is also provided with BEPS (Battery
Error Prevention System), which automatically
switches the meter off if the battery level is so
low to cause erroneous readings.
It is recommended to replace
immediately the battery.
Turn the meter off, remove the
battery compartment cover from
the rear of the meter and re-
place the rundown 9V battery
with a new one.
Install the battery while paying
attention to its polarity and re-
attach the cover.
WARRANTY:
This instrument is warranted against defects in materi-
als and manufacturing for a period of two years from
the date of purchase. Probe is warranted for 6 months.
The probes are warranted for a period of six months.
If during this period the repair or replacement of parts is
required, where the damage is not due to negligence or
erroneous operation by the user, please return the parts
to either distributor or our office and the repair will be
effected free of charge.
Damages due to accidents, misuse, tampering or lack
of prescribed maintenance are not covered.
NOTES:
If reading is over-range, the display will flash
the closest full scale value.
To clear a previous calibration, press “ON/
OFF” after entering the calibration mode. The
lower LCD will display ESC for 1 second and
the meter will return to normal measurement
mode. The “CAL” symbol on the LCD will
disappear and the meter is reset to the default
calibration.
OPTIONAL ACCESSORIES:
MA9004 pH 4.01 buffer solution, 230 ml bottle
MA9006 pH 6.86 buffer solution, 230 ml bottle
MA9007 pH 7.01 buffer solution, 230 ml bottle
MA9009 pH 9.18 buffer solution, 230 ml bottle
MA9010 pH 10.01 buffer solution, 230 ml bottle
MA9030 12.88 mS/cm cal.solution, 230 ml bottle
MA9015 Probe storage solution, 230 ml bottle
MA9016 General cleaning sol., 230 ml bottle
M10000B Rinse solution, 20 ml sachet (25 pcs.)
MA851D/1 Amplified pH/EC/TDS/Temperature
probe with DIN connector and 1 m
(3.3’) cable.
SPECIFICATIONS:
RANGE
0.00 to 14.00 pH
0.00 to 20.00 mS/cm
0.00 to 10.00 ppt
0.0 to 60.0ºC or 32.0 to 140.0°F
RESOLUTION
0.01 pH
0.1 mS/cm
0.01 ppt
0.1°C or 0.1°F
ACCURACY (@25°C) ±0.01pH
±2% FS (EC/TDS)
±0.5°C or ±1°F
TYPICAL EMC DEV. ±0.02 pH
±2% FS (EC/TDS)
±0.5°C or ±1°F
TEMPERATURE Automatic from 0 to 60ºC;
COMPENSATION ß adj. from 0.0 to 2.4%/°C
pH CALIBRATION 1 or 2-point
with automatic buffer recognition
EC CALIBRATION Automatic, 1 point
EC/TDS CONVERSION adj. from 0.45 to 1.00
FACTOR
PROBE (included) MA851D/1
amplified pH/EC/TDS/Temperature probe
with DIN connector and 1 m (3.3’) cable.
ENVIRONMENT 0 to 50°C (32 to 122°F);
100% RH max.
BATTERY TYPE 1 x 9V (6F22)
BATTERY LIFE approx. 300 hours
AUTO-OFF after 8 minutes
DIMENSIONS 200 x 85 x 50 mm
WEIGHT 260 g (with battery)
pH
EC
TDS
Temp.
pH
EC
TDS
Temp.