SPECIFICATIONS:
RANGE -2.00 to 16.00 pH
-5 to 70°C
RESOLUTION 0.01 pH / 0.1°C
ACCURACY (@25°C) ±0.02 pH / ±0.5°C
TYPICAL EMC DEV. ±0.02 pH / ±0.5°C
TEMPERATURE Automatic,
COMPENSATION 0 to 70°C
CALIBRATION Automatic, at 1 or 2 points
pH ELECTRODE MA911B/1 (included)
TEMPERATURE MA830R (included)
PROBE
ENVIRONMENT 0 to 50°C; RH 95% max.
BATTERY TYPE 1 x 9V alkaline (included)
BATTERY LIFE approx. 750 hours of use
AUTO-OFF after 8 minutes of non-use
DIMENSIONS 145 x 80 x 40 mm
WEIGHT 220 g (with battery)
Note: Pressing the CAL key again will
change the required buffer value,
making possible a single point cali-
bration at pH 4.01 or pH 10.01.
Wait till the hourglass symbol on the
display turns off and the “pH” tag stops
blinking, then
press the CFM key
to confirm the off-
set calibration. “pH
4.01” will flash on
the LCD.
Rinse the pH electrode
and the temperature
probe in the first pH4.01
beaker, then immerse
them into the second
pH4.01 beaker.
Follow the same procedure if using the
pH10.01 solution; select the correct
buffer by pressing the CAL key.
Wait till the hourglass symbol on the
LCD turns off, then press the CFM key
to confirm the slope calibration.
Notes:
To exit the calibration mode, turn the
meter off by pressing the ON/OFF key.
To perform a single point calibration
(offset) at pH7.01, press the ON/OFF
key when the second buffer is asked.
After battery replacement, the meter dis-
plays the “Clr” message and
recalibration is needed.
BATTERY REPLACEMENT:
When the battery becomes so weak that
no reliable measurement can be guaran-
teed, the meter will display “Eb” for a while
and then turns off.
It is recommended to replace immediately
the battery.
Battery replacement must only take place
in a non-hazardous area using a 9V alka-
line battery.
Turn the meter off, slide
off the battery compart-
ment cover at the rear of
the meter and replace
the 9V battery with a new
one.
Make sure the battery
contacts are tight and
secure, seat the battery
in its compartment and
insert back the cover.
OPTIONAL ACCESSORIES:
MA9004 pH4.01 buffer solution,
220 mL bottle
MA9007 pH7.01 buffer solution,
220 mL bottle
MA9010 pH10.01 buffer solution,
220 mL bottle
MA9015 Electrode storage solu-
tion, 220 mL bottle
MA9016 General cleaning solu-
tion, 220 mL bottle
M10000B Rinse solution, 20 ml
sachet (25 pcs.)
MA911B/1 pH electrode with BNC
connector and 1 m
cable
MA830R Temperature probe
MA950 Portable meter wall fix-
ing kit