Milwaukee
ISMIL102R1 01/2003
SPECIFICATIONS:
RANGE -2.00 to 16.00 pH
-5 to 70°C
RESOLUTION 0.01 pH / 0.1°C
ACCURACY (@25°C) ±0.02 pH / ±0.5°C
TYPICAL EMC DEV. ±0.02 pH / ±0.5°C
TEMPERATURE Automatic,
COMPENSATION 0 to 70°C
CALIBRATION Automatic, at 1 or 2 points
pH ELECTRODE MA911B/1 (included)
TEMPERATURE MA830R (included)
PROBE
ENVIRONMENT 0 to 50°C; RH 95% max.
BATTERY TYPE 1 x 9V alkaline (included)
BATTERY LIFE approx. 750 hours of use
AUTO-OFF after 8 minutes of non-use
DIMENSIONS 145 x 80 x 40 mm
WEIGHT 220 g (with battery)
Note: Pressing the CAL key again will
change the required buffer value,
making possible a single point cali-
bration at pH 4.01 or pH 10.01.
• Wait till the hourglass symbol on the
display turns off and the “pH” tag stops
blinking, then
press the CFM key
to confirm the off-
set calibration. “pH
4.01” will flash on
the LCD.
• Rinse the pH electrode
and the temperature
probe in the first pH4.01
beaker, then immerse
them into the second
pH4.01 beaker.
• Follow the same procedure if using the
pH10.01 solution; select the correct
buffer by pressing the CAL key.
• Wait till the hourglass symbol on the
LCD turns off, then press the CFM key
to confirm the slope calibration.
Notes:
• To exit the calibration mode, turn the
meter off by pressing the ON/OFF key.
• To perform a single point calibration
(offset) at pH7.01, press the ON/OFF
key when the second buffer is asked.
• After battery replacement, the meter dis-
plays the “Clr” message and
recalibration is needed.
BATTERY REPLACEMENT:
When the battery becomes so weak that
no reliable measurement can be guaran-
teed, the meter will display “Eb” for a while
and then turns off.
It is recommended to replace immediately
the battery.
Battery replacement must only take place
in a non-hazardous area using a 9V alka-
line battery.
Turn the meter off, slide
off the battery compart-
ment cover at the rear of
the meter and replace
the 9V battery with a new
one.
Make sure the battery
contacts are tight and
secure, seat the battery
in its compartment and
insert back the cover.
OPTIONAL ACCESSORIES:
MA9004 pH4.01 buffer solution,
220 mL bottle
MA9007 pH7.01 buffer solution,
220 mL bottle
MA9010 pH10.01 buffer solution,
220 mL bottle
MA9015 Electrode storage solu-
tion, 220 mL bottle
MA9016 General cleaning solu-
tion, 220 mL bottle
M10000B Rinse solution, 20 ml
sachet (25 pcs.)
MA911B/1 pH electrode with BNC
connector and 1 m
cable
MA830R Temperature probe
MA950 Portable meter wall fix-
ing kit