USER MANUAL
PORTABLE uP-BASED
pH METER
MODEL: SM102
Smart pH Meter
OPERATION:
The meter is supplied complete with a
9V battery. Slide off the battery compart-
ment cover on the back of the meter
and install the battery while paying at-
tention to its polarity.
Always remove the electrode protective
cap before taking any measurement. If
the electrode has been left dry, soak
the tip (bottom 2.5 cm) in rinse solution
(M10000B) for a few minutes to reacti-
vate it.
Connect the pH
electrode and the
temperature probe
to the sockets on
the meter’s top.
The temperature
probe can be used
independently to
take temperature
measurements, or
it can be used in
conjunction with
the pH electrode to utilize the meter’s
ATC capability.
Turn the instrument on by pressing the
ON/OFF key.
Before taking any mea-
surements, make sure
that the meter has al-
ready been calibrated.
Immerse the tip (2.5 cm)
of the pH electrode and
the temperature probe
into the sample and stir
gently.
Take the pH reading when the hour-
glass symbol stops blinking.
To display the mea-
sured temperature,
press and hold the
°C key.
When the key is released, the display
will return to the pH reading.
The HOLD function is activated by
keeping pressed the HOLD key.
The measured value is frozen on
the display and the “HOLD” tag
lights up.
After measurements, switch the
meter off and store the electrode
with a few drops of storage solu-
tion in the protective cap.
Error message on display:
CALIBRATION PROCEDURE:
A) Preparation:
Reference solutions:
1. pH 7.01 (MA 9007)
2. pH 4.01 (MA 9004) for measuring
in the acid range (pH lower than 7)
or pH 10.01 (MA 9010) for measur-
ing in the alkaline range (pH higher
than 7).
Use two beakers for each reference so-
lution. One beaker for rinsing the
electrode, the other for calibration. This
way contamination between solutions
during calibration is minimized.
B) Procedure:
Remove the electrode
protective cap and rinse
the tip with some
pH7.01 solution, then
immerse the electrode
and the temperature
probe into a pH7.01
buffer solution.
Press the CAL key
to enter the Cali-
bration mode. The
display flashes
“pH 7.01” and the
hourglass lights up.
Blinking full scale value: reading is
over range;
“Eb”: battery low voltage indication;
“Ec”: wrong buffer solution;
“Clr”: the calibration data have been
lost and recalibration is needed.